Skip to main content

Table 1 Bracket slot height in the analysed systems. In boldface the systems with a mean slot height out of ISO 27020:2019 tolerance limit

From: Compliance with the ISO 27020:2019 norm of a sample of currently available preadjusted Orthodontic bracket systems. Are the actual dimensions as expected?

Bracket system

N

Mean (± SD)

Min

Max

MD

ZW

p

SYNTHESIS

30

0.556(± 0.004)

0.554

0.559

-0.003

-4.616

< 0.001***

NU-EDGE

30

0.574(± 0.010)

0.559

0.591

0.015

-4.784

< 0.001***

DISCOVERY

30

0.570(± 0.010)

0.558

0.590

0.011

-4.476

< 0.001***

FLI TWIN

30

0.566(± 0.008)

0.558

0.590

0.007

-4.560

< 0.001***

MINI SPRINT

30

0.567(± 0.005)

0.558

0.577

0.008

-4.705

< 0.001***

OMNIARCH +

30

0.560(± 0.002)

0.555

0.567

0.001

-2.575

0.010**

SCAPE

30

0.562(± 0.006)

0.558

0.571

0.003

-3.564

< 0.001***

MINI-TWIN

30

0.554(± 0.005)

0.542

0.562

-0.005

-3.924

< 0.001***

JAZZ

30

0.548(± 0.009)

0.528

0.560

-0.011

-4.456

< 0.001***

MINI MASTER

30

0.548(± 0.005)

0.540

0.559

-0.011

-4.766

< 0.001***

MINIPREVAIL

30

0.560(± 0.004)

0.552

0.570

0.001

-1.763

0.078

VICTORY

30

0.560(± 0.001)

0.558

0.563

0.001

-3.953

< 0.001***

  1. SD Standard deviation; Min Minimum slot height; Max Maximum slot height; MD Mean difference between nominal and actual values; ZW Wilcoxon standardized rank; Wilcoxon test *p ≤ .05; **p < .01; ***p < .001