From: Incorrect measurements and misleading conclusions in the article “Comparison of the efficacy of tooth alignment among lingual and labial brackets: an in vitro study”

Fig. 1

a An identical set up as in the study of Alobeid et al.: Acrylic resin model (Palavit G 4004; Heraeus Kulzer, Hanau, Germany) was fabricated from a duplicate of a Frasaco model (Frasaco, Tettnang, Germany) of a normal maxillary arch. The upper-right, central incisor was removed. The model was bonded with conventional brackets with 0.022” slot size (GAC Twin, Dentsply Sirona, Charlotte, USA). A 0.014” Thermaloy-NiTi arch-wire (RMO, Denver, USA) was inserted. The stainless steel ligatures used were tied using a needle holder. The ligature was first tightened around the bracket wings and then loosened by one turn, to allow free movement of the arch-wire. b The reference pin was placed at a distance of 2 mm from the arch-wire. c The simulation was carried out at an ambient temperature of 36C. A horizontal displacement of 2 mm was simulated. At the end of the displacement, the wire was stuck, because of friction and binding, and did not move back at all (correction = 0%). Alobeid et al. reported a correction of 1.6 mm, equal to 82%