Skip to main content

Table 5 Axis angles errors (°) for the different surgical guide methods

From: Precision of fibula positioning guide in mandibular reconstruction with a fibula graft

Measurement line Fibula cutting guide alone (control group) Fibula cutting guide & fibula positioning guide (experimental group) P value
Anterior segment axis 9.9 ± 5.8 8.1 ± 5.3 0.375
Middle segment axis 9.5 ± 4.6 7.8 ± 4.1 0.285
Posterior segment axis 12.8 ± 8.1 9.1 ± 6.8 0.195
  1. Note: Values represent the mean ± standard deviation
  2. *P < 0.05