Table 4 Splint accuracy (Bland-Altman upper and lower limits) of 3D translational and rotational differences between the planned and post-operative images

From: Accuracy evaluation of CAD/CAM generated splints in orthognathic surgery: a cadaveric study

Translational difference (95 % CI) Rotational difference (95 % CI)
  Lower limit Upper limit   Lower limit Upper limit
Mediolateral −2.71 (−4.12 to −1.30) 1.83 (0.41 to 3.24) Pitch −9.33 (−14.16 to −4.49) 6.22 (1.38 to 11.06)
Anteroposterior −2.23 (−4.09 to −0.37) 3.74 (1.88 to 5.60) Roll −4.60 (−6.92 to −2.27) 2.87 (0.54 to 5.19)
Superoinferior −1.63 (−2.87 to −0.39) 2.35 (1.11 to 3.59) Yaw −1.51 (−2.81 to −0.22) 2.65 (1.35 to 3.94)