Skip to main content

Table 2 Peak skin and absorbed tissue doses for the investigated scan protocols

From: Measurement of skin dose from cone-beam computed tomography imaging

X-ray unit

kV

mAs

sFOV

Skin

Eye lens

Parotid

Submandibular

Thyroid

Dose (mGy)

Dose (mGy)

Dose (mGy)

Dose (mGy)

Dose (mGy)

Kodak 9500

86

108

15×9

3.58

0.42

2.88

3.04

0.32

Kodak 9500

120

108

20×18

4.15

3.55

3.57

3.17

0.45

iCAT Next Generation

120

18.54

16×13

1.44

0.92

1.15

1.21

0.24

iCAT Next Generation

120

20.27

16×13

2.51

1.60

2.05

1.79

0.25

iCAT Next Generation

120

18.54

23×17

2.08

1.41

1.93

1.79

0.25

iCAT Next Generation

120

37.07

23×17

2.60

1.75

2.11

2.23

0.39

Galileos Comfort

85

21

15×15

2.49

0.58

2.45

1.93

0.36

Galileos Comfort

85

42

15×15

4.62

0.94

4.46

3.61

0.46

ProMax panoramic

66

144

 

0.26

0.05

0.24

0.14

0.04

ProMax cephalometric

68

112.2

 

0.09

0.07

0.06

0.06

0.05