Comparative analysis of slot dimension in lingual bracket systems

Abstract

Background

Orthodontic treatment with fixed appliances requires - among others - the correct clinical expression of torque, which depends on the precise fitting of archwire and slot. Especially in the lingual technique torque problems become clinically more evident than in labial appliances also with respect to the vertical alignment of teeth due to different distances from the center of resistance. The purpose of the present study was to compare the preciseness of slot dimensions of different lingual bracket systems.

Methods

Three lingual bracket systems were included in the study (7th Generation and STb, Ormco, Glendora, CA, USA; Incognito, TOP-Service/3 M Unitek, Monrovia, CA, USA). Non destructive analysis of vertical slot dimensions was performed using precision pin gauges (Azurea, Belprahon, Switzerland) that were tapered in increments of 0.002 mm (0.00008 inch). The sizes of 240 incisor and canine brackets were measured per system (total: 720). Data were compared using one-way ANOVA. A p-value < 0.05 was considered statistically significant.

Results

Average slot dimensions were 0.467 mm ± 0.007 mm (0.0184 inch ± 0.0003 inch) for the 7th Generation bracket system, 0.466 mm ± 0.004 mm (0.0183 inch ± 0.0001) inch for the STb bracket system and 0.459 mm ± 0.004 mm (0.0181 inch ± 0.0001) inch for the Incognito bracket system. Differences between systems were statistically significant (p < 0.05).

Conclusions

The analyzed bracket systems for lingual treatment exhibited significant differences in slot dimension that will clinically result in torque play. These aspects must be considered in lingual orthodontic treatment.

Peer Review reports

Background

In recent years, an increased number of adult orthodontic patients have sought the correction of malaligned teeth. These patients are frequently focused on dental esthetics in the anterior region. For socio-cultural reasons, adult patients favor invisible orthodontic treatment over eye-catching treatment by buccal appliances. Hence, fixed lingual bracket systems have been developed over the last 30 years [14]. Due to the large bucco-lingual dimension of these prefabricated bracket systems, these systems were reported to cause clinical problems, such as speech dysfunction as a result of restricted functional space for the tongue, oral discomfort due to injury or irritation of the tongue, and restriction of mastication [57]. The implementation of customized lingual brackets and computerized archwire fabrication resulted in a decrease of subjective impairments [810]. Thanks to clinical simplification in using customized appliances, the indication for lingual orthodontics has been extended to adolescents [11].

Besides visibility of the appliance during orthodontic treatment, the labio-lingual inclination of maxillary and mandibular incisors and canines is considered by patients and orthodontists to be an important determinant in providing esthetic outcome after orthodontic treatment. Furthermore, the correct inclination of the anterior teeth is essential in providing good occlusion in anterior and posterior regions, and is basically dependent on the correct expression of torque. This torque expression can be achieved by using slot-filling archwires or twisting and inserting undersized archwires. Nevertheless, various factors affect the correct torque expression of a preadjusted appliance [12]. These factors include material properties such as hardness and modulus of elasticity of the archwire and bracket, variations of manufacturing processes including milling and casting of brackets, as well as clinical procedures like mode of ligation [13, 14]. Furthermore, a patient's individual tooth morphology also influences the clinical outcome of torque [15, 16]. In the literature, it was shown that oversized slots lead to a clinically relevant torque loss [17]. Due to the longer lever arm in lingual orthodontics, slot precision must be considered as a key factor that influences tooth position not only in the labio-lingual but also in the vertical dimension [18] (see Figure 1).

Figure 1
figure 1

Effects of torque loss. Comparison of torque loss (-15 degrees) in lingual and buccal technique with respect to vertical side effects (CR = Centre of Rotation).

Therefore, the aim of this study was to evaluate the slot precision of frequently used lingual appliances.

Methods

Three lingual bracket systems were included in the study: 7th Generation (Ormco, Glendora, CA, USA), STb (Ormco) and Incognito (TOP Service/3 M Unitek, Monrovia, CA, USA). All brackets were manufactured with a slot dimension of 0.457 mm (0.018) inch. For each system, the sizes of 240 incisor and canine bracket slots were measured (a total of 720). Non-destructive analysis of 0.018 inch slot dimensions was performed by one operator using precision pin gauges (Azurea, Belprahon, Switzerland). Pin gauges were tapered in increments of 0.002 mm (0.00008 inch), Figure 2.

Figure 2
figure 2

Pin gauges. Precision pin gauges tapered in increments of 0.002 mm (0.00008 inch).

Measurements of slot precision were performed starting with the insertion of the smallest pin gauge with a dimension of 0.456 mm (0.01795 inch), Figure 3. Consecutive insertion of pin gauges was continued until the bracket was attached to the pin gauge by friction (Figure 4). The value of the pin gauge one increment below the finally inserted one was documented as slot size.

Figure 3
figure 3

Start of measurement. Start of measurement of slot precision by insertion of the smallest pin gauge with a dimension of 0.456 mm (0.01795 Inch).

Figure 4
figure 4

End of measurement. Measurement of slot precision ended when the bracket was attached to the pin gauge by friction.

Power and sample sizes were calculated using nQuery Advisor 5.0 (Statistical Solutions, Saugas, Massachusetts, USA). Power calculation revealed that the ANOVA F-test with a sample size of 240 would have a 90% power to detect a difference in means of +0.00254 mm (+0.0001 inch) and -0.00254 mm (-0.0001) inch to a reference group, assuming a within-group standard deviation of 0.01524 mm (0.0006 inch). Documentation and evaluation of the data was performed using the Statistical Package for the Social Sciences, Version 17.0 for Windows (SPSS, Chicago, Illinois, USA). The Kolmogorov-Smirnov test was applied to test for normal distribution. As data were distributed normally, they were compared using one-way ANOVA. Post-hoc testing was performed with the Fisher's LSD test. All tests were performed two-tailed with a significance level of α = 0.05.

Results

Results of experimental measurements are presented as a boxplot in Figure 5 and in Table 1. Average slot dimensions were 0.467 mm ± 0.007 mm (0.0184 inch ± 0.0003 inch) for the 7th Generation bracket system, 0.466 mm ± 0.004 mm (0.0183 inch ± 0.0001 inch) for the STb bracket system and 0.459 mm ± 0.004 mm (0.0181 inch ± 0.0001 inch) for the Incognito system. The calculated medians for the included lingual bracket systems were 0.4660 mm (0.0184 inch) for the 7th Generation and the STb bracket and 0.4580 mm (0.0180 inch) for the Incognito bracket. Statistical analysis revealed significant differences in slot precision between all analyzed bracket systems, with the highest precision for the Incognito bracket and the lowest for the 7th Generation bracket (p < 0.001).

Table 1 Means, minimum and maximum values of slot precision (mm) with respect to bracket location
Figure 5
figure 5

Slot precision. Results of determined slot precision (* = p < 0.001).

Discussion

The clinical expression of torque can be achieved by inserting archwires with gradually increased dimensions. However, due to undersized archwire dimensions, edge bevelling and oversized slot dimensions, a significant play of the archwire in the slot can be observed [1921]. These factors result in a clinically relevant torque loss.

Therefore, the literature describes various methods for the determination of slot precision. For example, a metallurgic method was used to measure bracket slot height in stainless steel brackets after preparation of cross-sectional cuts through the bracket slot [22]. Afterwards, the slot height was measured to the nearest 0.01 mm with a Zeiss Axioscope. This method was characterized by destructive analysis of brackets, considerable technical effort, and low accuracy. Consequently, methods were improved by estimating effective slot height by using a formula that describes the relationship between bracket slot height, wire dimensions, wire edge bevel and torsional play [17]. This method has the advantage of high precision and non-destructive analysis, but again requires great technical effort. To allow determination of slot precision in large samples with a high degree of accuracy, measurements in the present study were performed by use of high precision pin gauges. The applied increments of 0.002 mm (0.00008 inch) can be considered as very finely graduated. The application of manual measurement furthermore allowed non-destructive analysis, and was therefore associated with low costs.

The standards for the precision of orthodontic attachments in Germany are defined in DIN 13971-2. According to this standard, a 0.0018 inch slot should have a slot width between 0.0018 inch (0.46 mm) and 0.00197 inch (0.50 mm) [23]. In the present investigation, all lingual bracket systems fulfilled these requirements, with measured average slot precisions between 0.0181 inch for the Incognito bracket and 0.0184 inch for the 7th Generation bracket.

However, significant differences in slot precision were found between all investigated bracket systems. The extremely high accuracy of the Incognito bracket can be explained by the manufacturing process. All Incognito brackets undergo subsequent manual measurement of slot precision [8]. Therefore, the tolerance range of slot dimension guaranteed by the manufacturer is 0.0180 inch to 0.0183 inch.

Although the other two brackets fulfilled the DIN standard, the clinical implications of torque play must be carefully considered. The correct buccolingual inclination of maxillary teeth is particularly critical in establishing an esthetic smile line, proper anterior guidance, and a Class I canine and molar relationship [24]. One study showed that an additional anterior inclination of 5° generates 1 mm of additional arch length [25]. Consequently, the correct clinical expression of torque can be regarded as a key factor for successful orthodontic treatment, but which is mechanically jeopardized in lingual orthodontics by the longer lever arm [18]. As the Incognito bracket is a customized bracket, it has a shorter lever arm than the other two lingual bracket systems. This factor must be regarded in addition to the significant differences in slot precision when determining the clinical performance of lingual bracket systems with respect to the clinical expression of torque.

Conclusions

In the present study, all investigated lingual bracket systems fulfilled the DIN 13971-2 standard with average slot sizes of 0.467 mm (0.0184 inch) for the 7th Generation bracket system, 0.466 mm (0.0183 inch) for the STb bracket system and 0.459 mm (0.0181 inch) for the Incognito system. However, slot precision differed statistically significantly between all evaluated bracket systems.

References

  1. Fujita K: New orthodontic treatment with lingual bracket mushroom arch wire appliance. Am J Orthod. 1979, 76: 657-675. 10.1016/0002-9416(79)90211-2.

    Article  CAS  PubMed  Google Scholar 

  2. Alexander CM, Alexander RG, Gorman JC, Hilgers JJ, Kurz C, Scholz RP, Smith JR: Lingual orthodontics. A status report. J Clin Orthod. 1982, 16: 255-262.

    CAS  PubMed  Google Scholar 

  3. Kurz C, Swartz ML, Andreiko C: Lingual orthodontics: a status report. Part 2: Research and development. J Clin Orthod. 1982, 16: 735-740.

    CAS  PubMed  Google Scholar 

  4. Takemoto K, Scuzzo G: The straight-wire concept in lingual orthodontics. J Clin Orthod. 2001, 35: 46-52.

    CAS  PubMed  Google Scholar 

  5. Fritz U, Diedrich P, Wiechmann D: Lingual technique--patients' characteristics, motivation and acceptance. Interpretation of a retrospective survey. J Orofac Orthop. 2002, 63: 227-233. 10.1007/s00056-002-0124-3.

    Article  PubMed  Google Scholar 

  6. Fillion D: Improving patient comfort with lingual brackets. J Clin Orthod. 1997, 31: 689-694.

    CAS  PubMed  Google Scholar 

  7. Fujita K: Multilingual-bracket and mushroom arch wire technique. A clinical report. Am J Orthod. 1982, 82: 120-140. 10.1016/0002-9416(82)90491-2.

    Article  CAS  PubMed  Google Scholar 

  8. Wiechmann D: A new bracket system for lingual orthodontic treatment. Part 2: First clinical experiences and further development. J Orofac Orthop. 2003, 64: 372-388. 10.1007/s00056-003-0325-4.

    Article  PubMed  Google Scholar 

  9. Stamm T, Hohoff A, Ehmer U: A subjective comparison of two lingual bracket systems. Eur J Orthod. 2005, 27: 420-426. 10.1093/ejo/cji034.

    Article  PubMed  Google Scholar 

  10. Wiechmann D: A new bracket system for lingual orthodontic treatment. Part 1: Theoretical background and development. J Orofac Orthop. 2002, 63: 234-245. 10.1007/s00056-002-0211-5.

    Article  PubMed  Google Scholar 

  11. Wiechmann D, Schwestka-Polly R, Hohoff A: Herbst appliance in lingual orthodontics. Am J Orthod Dentofacial Orthop. 2008, 134: 439-446. 10.1016/j.ajodo.2007.09.015.

    Article  PubMed  Google Scholar 

  12. Gioka C, Eliades T: Materials-induced variation in the torque expression of preadjusted appliances. Am J Orthod Dentofacial Orthop. 2004, 125: 323-328. 10.1016/j.ajodo.2003.02.007.

    Article  PubMed  Google Scholar 

  13. Meling TR, Odegaard J: On the variability of cross-sectional dimensions and torsional properties of rectangular nickel-titanium arch wires. Am J Orthod Dentofacial Orthop. 1998, 113: 546-557. 10.1016/S0889-5406(98)70266-5.

    Article  CAS  PubMed  Google Scholar 

  14. Morina E, Eliades T, Pandis N, Jager A, Bourauel C: Torque expression of self-ligating brackets compared with conventional metallic, ceramic, and plastic brackets. Eur J Orthod. 2008, 30: 233-238. 10.1093/ejo/cjn005.

    Article  PubMed  Google Scholar 

  15. Germane N, Bentley BE, Isaacson RJ: Three biologic variables modifying faciolingual tooth angulation by straight-wire appliances. Am J Orthod Dentofacial Orthop. 1989, 96: 312-319. 10.1016/0889-5406(89)90350-8.

    Article  CAS  PubMed  Google Scholar 

  16. Miethke RR, Melsen B: Effect of variation in tooth morphology and bracket position on first and third order correction with preadjusted appliances. Am J Orthod Dentofacial Orthop. 1999, 116: 329-335. 10.1016/S0889-5406(99)70246-5.

    Article  CAS  PubMed  Google Scholar 

  17. Meling TR, Odegaard J, Seqner D: On bracket slot height: a methodologic study. Am J Orthod Dentofacial Orthop. 1998, 113: 387-393.

    CAS  PubMed  Google Scholar 

  18. Stamm T, Wiechmann D, Heinecken A, Ehmer U: Relation between second and third order problems in lingual orthodontic treatment. J Ling Orthod. 2000, 3: 5-11.

    Google Scholar 

  19. Meling TR, Odegaard J, Meling EO: On mechanical properties of square and rectangular stainless steel wires tested in torsion. Am J Orthod Dentofacial Orthop. 1997, 111: 310-320. 10.1016/S0889-5406(97)70190-2.

    Article  CAS  PubMed  Google Scholar 

  20. Cash AC, Good SA, Curtis RV, McDonald F: An evaluation of slot size in orthodontic brackets--are standards as expected?. Angle Orthod. 2004, 74: 450-453.

    CAS  PubMed  Google Scholar 

  21. Sebanc J, Brantley WA, Pincsak JJ, Conover JP: Variability of effective root torque as a function of edge bevel on orthodontic arch wires. Am J Orthod. 1984, 86: 43-51. 10.1016/0002-9416(84)90275-6.

    Article  CAS  PubMed  Google Scholar 

  22. Meling T, Odegaard J, Holthe K, Meling EO, Segner D: A formula for the displacement of an arch wire when subjected to a second-order couple. Am J Orthod Dentofacial Orthop. 1998, 113: 632-640. 10.1016/S0889-5406(98)70223-9.

    Article  CAS  PubMed  Google Scholar 

  23. Sernetz F: Standardization of orthodontic products--does it make sense?. J Orofac Orthop. 2005, 66: 307-318. 10.1007/s00056-005-0503-7.

    Article  PubMed  Google Scholar 

  24. Badawi HM, Toogood RW, Carey JP, Heo G, Major PW: Torque expression of self-ligating brackets. Am J Orthod Dentofacial Orthop. 2008, 133: 721-728. 10.1016/j.ajodo.2006.01.051.

    Article  PubMed  Google Scholar 

  25. O'Higgins EA, Kirschen RH, Lee RT: The influence of maxillary incisor inclination on arch length. Br J Orthod. 1999, 26: 97-102. 10.1093/ortho/26.2.97.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

We would like to thank Prof. Dr. em. H. Hecker, Institute of Biostatistics, Hannover Medical School for assistance with statistical analysis. Furthermore we thank Mrs. S. Robra for the illustrations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Anton Demling.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

AD performed measurement of slot dimensions, statistics and wrote major parts of the manuscript. MPD assisted in performance of statistics and wrote parts of the manuscripts. RSP conceived the study design and participated in preparation of the manuscript including interpretation of data. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Demling, A., Dittmer, M.P. & Schwestka-Polly, R. Comparative analysis of slot dimension in lingual bracket systems. Head Face Med 5, 27 (2009). https://doi.org/10.1186/1746-160X-5-27

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1746-160X-5-27

Keywords