Table 4 Overview on CTDIVol and DLP values for each scanning protocol for trauma diagnostic

From: Dose development in sinonasal imaging over the last decade – a retrospective patient study

 

CTDIVol

 

DLP

Scanning protocol

Mean

(mGy)

Reduction in % compared to DS 64 / FBP

 

Mean ± Std. Dev.

(mGy x cm)

Reduction in % compared to DS 64 / FBP

DS 64 / FBP

14.9

-

 

232

±

43

-

DS 64 / 1st Gen.

13.7

9

 

218

±

78

6

DS 64 / 2nd Gen.

10.6

29

 

168

±

36

27

SS 128 / FBP

10.9

27

 

154

±

27

34

SS 128 / 2nd Gen.

8.7

42

 

134

±

29

42

DS 192 / 3rd Gen. + ATCM

11.9

20

 

199

±

47

14

DS 192 / 3rd Gen. - ATCM

8.8

41

 

140

±

25

40

DS 192 / 3rd Gen. 2022

5.0

66

 

76

±

11

67

  1. Mean value (± standard deviation) of CTDIVol and DLP values of protocols used for trauma assessment in the past 12 years. Additionally, the dose reduction in percent per scan protocol compared the DS 64 / FBP protocol is shown for the CTDIVol and DLP values. ATCM = automatic tube current modulation, CBCT = cone beam CT, CT = computed tomography, CTDIVol = computed tomography dose index, DLP = dose length product, FBP = filtered back projection, Gen. = generation.