Skip to main content

Table 1 Overview on image acquisition and image reconstruction parameters

From: Dose development in sinonasal imaging over the last decade – a retrospective patient study

 

Image acquisition

 

Image reconstruction

CT Scanner

Detector

Acquisition year

Tin filtration

Tube voltage (kV)

Tube current (mAs)

Collimation (mm) x number of slices

Pitch

Rotation time (1/s)

Automatic tube current modulation

 

Reconstruction method

Slice thickness (mm)

Increment

Kernel

Chronic sinusitis

Somatom Definition AS 64 (DS 2 × 64)

Ultra fast ceramic

2010–2011

-

120

25

0.6 × 32

1

1

-

 

FBP

1

1

H60s

2012–2014

-

120

25

0.6 × 32

1

1

-

 

IRIS

1

1

J70h

2015–2016

-

120

20

0.6 × 32

1

1

-

 

SAFIRE

1

1

H50s

Somatom Definition AS + 128 (SS 128)

Ultra fast ceramic

2011–2013

-

100

35

0.6 × 64

0.8

1

-

 

FBP

1

1

H60s

2014–2022

-

100

25

0.6 × 64

0.8

1

-

 

SAFIRE

1

1

J70h

Somatom Force

(DS 2 × 192)

Stellar Infinity

2016

X

100

Ref. 90

0.6 × 96

0.8

0.5

X

 

ADMIRE

2

1

Hr64d

2016–2019

X

100

50

0.6 × 96

0.6

0.5

-

 

ADMIRE

1

1

Hr64d

2020–2022

X

100

125

0.6 × 192

0.8

1

-

 

ADMIRE

1

1

Hr64d

3D Accuitomo 170

 

2015–2017

-

90

5–9

-

-

9–17

-

 

-

0.5

-

-

Preoperative

Somatom Definition AS 64 (DS 2 × 64)

Ultra fast ceramic

2010–2011

-

120

30

0.6 × 32

0.9

1

-

 

FBP

1

1

H60s

2012–2014

-

120

30

0.6 × 32

0.8

1

-

 

IRIS

1

1

J70h

2015–2016

-

120

30

0.6 × 32

0.9

1

-

 

SAFIRE

1

1

H70h

Somatom Definition AS + 128 (SS 128)

Ultra fast ceramic

2011–2013

-

100

30

0.6 × 64

0.8

1

-

 

FBP

1

1

H60s

2014–2022

-

100

25

0.6 × 64

0.8

1

-

 

SAFIRE

1

1

J70h

Somatom Force

(DS 2 × 192)

Stellar Infinity

2016

-

120

Ref. 63

0.6 × 96

0.8

1

X

 

ADMIRE

1

0.75

Hr64h

2016–2019

-

100

25

0.6 × 96

0.8

1

-

 

ADMIRE

1

0.75

Hr64h

2020–2022

X

100

Ref. 250

0.6 × 192

0.8

1

X

 

ADMIRE

1

1

Hr64h

Posttraumatic

Somatom Definition AS 64 (DS 2 × 64)

Ultra fast ceramic

2010–2011

-

120

100

0.6 × 32

1

1

-

 

FBP

1

1

H60s

2012–2014

-

120

100

0.6 × 32

1

1

-

 

IRIS

1

1

J70h

2015–2016

-

120

70

0.6 × 32

1

1

-

 

SAFIRE

1

1

H70h

Somatom Definition AS + 128 (SS 128)

Ultra fast ceramic

2011–2013

-

100

125

0.6 × 64

0.8

1

-

 

FBP

1

1

H60s

2014–2022

-

100

100

0.6 × 64

0.8

1

-

 

SAFIRE

1

1

J70h

Somatom Force

(DS 2 × 192)

Stellar Infinity

2016

-

120

Ref. 88

0.6 × 96

0.8

1

X

 

ADMIRE

2

1

Hr64h

2016–2019

-

100

100

0.6 × 96

0.8

1

-

 

ADMIRE

1

1

Hr64h

2020–2022

X

100

600

0.6 × 192

0.8

1

-

 

ADMIRE

1

1

Hr64h

  1. Overview on technical specification of the three used CT scanners (DS 2 × 64, SS 128, DS 2 × 192) and of the CBCT (3D Accuitomo 170), the image acquisition parameters and image reconstruction parameters of the protocols used in the past 12 years. CT = computed tomography, DS = dual source, FBP = filtered back projection, SS = single source.